Loadstar Sensors DS-3000 Digital Load Cell Display & Controller User Manual

Page 30

Advertising
background image

DS-3000U

DS3000 User Guide

Loadstar Sensors, Inc.

Page: 30 / 48

033-01419 Rev. 4

7. Select

pnt1rb for pnt1 using the enter key to select.

8. Press

down

to trgr 2, press enter and a new numeric field appears. Enter 2300 Lbs. Press

enter

when completed.

9. Select

pnt2ra for pnt2 using the enter key to select. This will reset the output, and terminate

the sequence, every time the weight rises above 2300 Lbs. The process cannot be restarted
until the remote input is triggered again…

10. Press down to sen1 corresponding to sensor 1. Press enter until

S

is displayed. This

means that the trigger condition will include the sum of sensor 1.

11. Press down and repeat for SEN2, SEN3 and SEN4. This way the sum of all four sensors will be

used to determine the trigger criteria.

12. Finally, press down to duratn, press enter and make sure it is set to contin. This means the

output will remain active during the whole time nly pulse for this interval when either the
minimum or maximum condition occurs.

13. Press enter to leave this field

14. Now exit programming mode: Press mode until the run entry

15. Then press enter to return to run mode.

16. Once in run mode, the program is now fully functional and operating. Try it out!

9.2 Empty Tank Warning

This application uses a scale platform, made with 4 sensors mounted on the corners of a square slab of
Aluminum, with a tank of liquid, in this example, a nitrogen Dewar atop the platform. Liquid Nitrogen, over
time is used up and the tank slowly loses weight. At some point the weight is reduced to the point where
the Dewar needs to be replaced- Since the replacement process can take a day or more, some advanced
warning is appropriate.

Therefore this application has two lights, yellow, which indicates the Dewar should be scheduled for
replacement, and red, which signals the Dewar is empty, and should be replaced. With both lights off, there
is nothing to do.

Table 1: Dewar Replacement Chart

Relay 1

N.O.

N.C.

N.C

Relay 2

N.O.

N.O.

N.C

LAMP

RED

Tank is Empty

YELLOW

Reorder Tank

NONE

0 Lbs

40 Lbs

70 LBS … 250 Lbs

9.2.1

Application Requirements:

1. 4x sensors each with 250 Lb load capacity each = 1000 Lb total scale load capacity

2. Assume this Dewar has a full weight of 250 Lbs, and an empty weight of 40 lbs. If typical usage

is 15 Lbs/day, and it takes 2 days to receive a shipment, then the yellow light should activate at
30+40=70 lbs, and the red light at 40 lbs (tank empty).

3. Let Sensor 1 – Sensor 4 correspond to the four load-cells, mounted to the measurement plate.

Each is connected to one of the four DS-3000 module inputs.