MTS SWIFT 40 Sensor User Manual
Page 21
Construction
SWIFT 40 Sensors
Hardware Overview
21
Transducer Interface
(TI)
The TI conditions the power supply, and uses previously stored calibration values
to convert the eight bridge outputs and the encoder signal to six non-rotating
analog outputs (Fx, Fy, Fz, Mx, My, Mz) plus an angle output. The force and
moment outputs have a value of 10 V full scale, unless a different full-scale
output is requested by a customer. The angle output is a 0–5 V sawtooth output.
Additional
components
Additional components that are supplied with your SWIFT sensor include shunt
and transducer data cables, TI power cable, a SWIFT Transducer Interface
Utilities disk, and the calibration file. MTS can also provide a 12 V DC power
converter for use in the test laboratory.