SWIFT 40 Sensors

52

TISHUNT - Transducer Interface Shunt

Software Utilities

This option asks if you want to save the data in a file. Answer Y or N, and if Y,
then supply the file name when prompted. The following is typical of what is
displayed when this selection is made:

Note

If you save to a file, the file is created in the same directory as the tishunt
program. The file name must contain less than 80 characters.

Option 2

Use this to set the shunt tolerance. When selected, the following is displayed:

Option 3

Use this option to restore shunt settings from a file. The file format is the same as
that used in the tixfer program.

Reading Shunt parameters from TI... OK

Shunt Delta Tol: 10.000%

FX1 Ref:

3.935

Measured:

3.934

Error Status:

0

FX2 Ref:

3.932

Measured:

3.932

Error Status:

0

FY1 Ref:

3.937

Measured:

3.930

Error Status:

0

FY2 Ref:

3.929

Measured:

3.930

Error Status:

0

FY3 Ref:

3.932

Measured:

3.932

Error Status:

0

FY4 Ref:

3.931

Measured:

3.932

Error Status:

0

FZ1 Ref:

3.930

Measured:

3.935

Error Status:

0

FZ2 Ref:

3.932

Measured:

3.934

Error Status:

0

Save to a file (y/n)?

The Shunt Tolerance specifies how much the measured shunt deltas may deviate

from the Ref settings, and still be considered good.

The value is entered as a percentage, where "1.0" would mean 1.0% of the Ref

settings.

For example: if the tolerance is 1%, and the Ref setting is 8.0, then the

measured delta must be between 7.92 and 8.08 to be considered good.

It is not necessary to reboot the TI when this value is changed, but a Shunt

Calibration must be executed to update the Shunt Error Status words.

Also note that a parameter download will overwrite this value.

Enter tolerance (in percent):