MTS SWIFT 40 Sensor User Manual

Page 36

Advertising
background image

SWIFT 40 Sensors

36

Transducer Interface

Hardware Overview

When a system zero is initiated, the associated LED is lit. After successful
completion of the system zero, the LED turns off. If there is a failure during the
system zero, the LED will flash at approximately a 1 Hz rate.

Angle Zero and Bridge Zero Switch Functions

Z

ERO

A

LGORITHM

W

HEN

TO

U

SE

A

NGLE

Z

ERO

B

RIDGE

Z

ERO

0

(Preferred method
for non-spinning
applications)

Use this algorithm for non-
spinning (fixed)
applications.

The Angle Zero switch is
non-functional.

Press the Bridge Zero
switch to measure the static
transducer bridge offsets.

1

Use this algorithm for
spinning applications.

You will need to
mechanically level the
SWIFT sensor.

Press the Angle Zero switch
to read the current angle and
use the value to set the angle
sum to 0.0°.

Press the Bridge Zero
switch to measure the static
transducer bridge offsets.

3

Use this algorithm for
spinning applications.

You do not need to
mechanically level the
SWIFT sensor.

Both the Angle Zero and Bridge Zero switches are
functional. This algorithm will perform both angle zero and
bridge zero processes regardless of which button you press.

4

(Preferred method
for spinning
applications)

Use this algorithm for
spinning applications.

You will need to
mechanically level the
SWIFT sensor.

Both the Angle Zero and Bridge Zero switches are
functional. This algorithm will perform both angle zero and
bridge zero processes.