Option 4 – MTS SWIFT 40 Sensor User Manual

Page 53

Advertising
background image

TISHUNT - Transducer Interface Shunt

SWIFT 40 Sensors

Software Utilities

53

Option 4

Use this option to apply a shunt sequentially, reading the eight bridges and six
outputs of the Transducer Interface. Note that this option does not update the
shunt measured values or error status. A reading is first made with no shunts
applied (displayed as the zeroes data). The remainder of the readings are
displayed as delta values from the original readings.

Enter Filename to save data to (the file will be appended):
Enter Description:
Reading bridges & outputs unshunted

FX1

FX

FX2

FY

FY1

FZ

FY2

MX

FY3

MY

FY4

MZ

FZ1

FZ2

zeroes:

bridges:

0.000

0.002

0.000

0.000

0.001

0.000

0.001

0.000

outputs:

-0.003

-0.006

0.001

0.000

-0.002

0.000

deltas:

Shunting FX1

bridges:

-3.933

0.000

-0.001

0.002

-0.001

0.001

0.000

0.000

outputs:

-9.014

0.031

-0.031

-0.001

-3.246

-0.188

Shunting FX2

bridges:

-0.002

-3.929

0.000

0.001

0.000

0.001

0.001

0.001

outputs:

-9.198

-0.049

0.026

-0.011

3.259

-0.072

Shunting FY1

bridges:

-0.001

0.001

-3.931

0.001

-0.001

0.000

-0.001

0.001

outputs:

0.029

-8.493

-0.145

-3.695

0.024

0.002

Shunting FY2

bridges:

0.000

-0.001

-0.001

-3.930

0.000

0.000

0.000

0.001

outputs:

0.175

-8.322

0.031

0.021

-0.014

-3.647

Shunting FY3

bridges:

-0.001

0.001

-0.001

0.000

-3.931

0.001

0.000

-0.001

outputs:

0.005

-8.538

0.163

3.692

-0.016

-0.003

Shunting FY4

bridges:

-0.002

0.001

0.000

0.000

0.001

-3.932

-0.001

0.000

outputs:

-0.150

-8.271

-0.018

-0.021

0.021

3.712

Shunting FZ1

bridges:

0.000

0.001

0.000

0.000

-0.001

0.002

-3.922

-0.001

outputs:

0.105

-0.010

-7.482

0.129

-2.892

-0.052

Shunting FZ2

bridges:

0.000

-0.001

0.000

0.001

0.000

0.001

-0.001

-3.931

outputs:

-0.052

-0.081

-7.021

0.114

2.749

0.049