64 • Last Calibration Window

Last Calibration Window

You can access data on the most recent successful calibration and bump
test for each instrument that has been downloaded in the Last Calibration
Window. You can view, print (calibration date/bump test date only), and
delete this data in the Last Calibration Window.

NOTE:

If the GX-2012 Data Logger Management Program database is
on the user’s computer (default setting), data will only appear for
instruments downloaded to that computer. If the database is on a
network, data will appear for all instruments downloaded to that
database. See “Accessing a Common Database” on page 14 for
more instruction.

Effects of BUMP DSP Parameter Setting on
Appearance of Last Calibration Window

The setting of the BUMP DSP instrument parameter has an effect on the
appearance of the Last Calibration Window. This instrument parameter can
be set in the instrument’s Maintenance Mode. The factory setting is Off.

If the BUMP DSP parameter is set to Off for every instrument that has been
downloaded to the GX-2012 Data Logger Management Program using the
Complete Download or Instrument Information download buttons, the Last
Calibration Window will appear as shown in the screen below. The screen
is shown with the Need Calibration option selected. The Bump Test portion
of this window is not shown.