70 • Last Calibration Window

Calibration Record/Bump Test Record View Option

Selecting the Calibration Record/Bump Test Record view option shows
detailed calibration or bump test information for each instrument’s most
recent successful calibration or bump test.

The fields on this screen include SerialNo, UserID, and StationID just as in
the Need Calibration/Need Bump Test and Calibration Date/Bump Test
Date screens. The Calibration Record fields also include the following:

Gas — Lists the target gas for which the Before, After, and A.Cal
readings are displayed.

Before — shows the settings prior to calibration.

After — shows the settings after calibration.

A. Cal. — lists the auto-calibration setting for each channel of the
instrument. If an instrument passes its calibration, the “After” column
should match the “A. Cal.” column. If the instrument fails calibration
on any of its channels, those channels will retain the previous
calibration information.

Figure 70: Last Calibration Window: Calibration Record View Option