Set Window • 73

NOTE:

Deleting instrument data from the folder list along the left side of
the Data screen will not delete calibration data from the Last
Calibration screen. Data in the Last Calibration screen must be
deleted manually.

Set Window

Changing Instrument Parameters

To make changes to parameters stored in the instrument, use the
GX-2012’s Data Logger Management Program’s Set Window. Follow the
steps below to make these changes.

1. Launch the GX-2012 Data Logger Management Program.

2. If the Automatic Download selection box is selected, deselect it.

3. Connect an instrument to the program. See “Downloading Data from

an Instrument” on page 17, instruction 1 through instruction 3, to
connect an instrument to the program.

4. Once the connection is made, the Complete Download, Instrument

Information, and Power Off download commands will be selectable.
Click the Instrument Information download command to retrieve the
instrument information from the instrument. If you wish to download
data before making changes, click Complete Download instead of
Instrument Information.