68 • Last Calibration Window

was calibrated or bump tested for each of the target gases using the
MM/DD/YY format and 24-hour military standard time.

Last Downloaded — shows when the last download took place for a
unit using the MM/DD/YY format and 24-hour military time standard.

The instruments that are due for calibration or bump testing (in the case of
the Need Calibration/Need Bump Test view option, that will be all of them),
will have their last calibration date or last bump test date highlighted in red.
The instruments that have not been downloaded for more than 90 days will
have their last download date highlighted in purple.

To print a list of the instruments shown in the Need Calibration/Need Bump
Test view option along with their user ID and last calibration date or last
bump test date, click the Print button. A Printer List dialog box will appear.

Select a printer and click the OK button to print the instrument list.