Bump Test Extension Module Chapter 2

See

Bump Test Parameter Descriptions on page 17

for complete

descriptions of the parameters.

5.

Configure the Bump Test setup parameters.

Table 2 - Bump Test Parameter Descriptions

Parameter

Description

Value

Num. Channels

Number of Channels you are using
in the Bump Test.

1 (X)

2 (X and Y)

3 (X, Y, and Z)

4 (X, Y, Z, R)

Sensor Type

Specifies the type of sensor used
(typically Accel G). Note that the
specified sensor type determines
available options and engineering
units for subsequent setup fields.

Accel (G) (default)

Accel m/s

2

Vel IPS

Vel mm/s

Disp um

Disp mil

Volts AC

Volts DC

Sensor Sensitivity

Enter transducer sensitivity in
millivolts (mV) per Engineering
Unit (EU) (typically 100 mV/g).

mV/EU

Range (EU) (Full Scale)

Define the spectrum’s maximum
full-scale amplitude value.

1 (default), 2, 5, 10, 20, 50, 100,
200, 500, 1000

Typically 20 G for Accel

X-axis units

Specify frequency units for graphic
displays.

Hz (default)

CPM

Filter

Displays high-pass filter settings.

2 Hz (default)

Off, 0.36 Hz, 1.1 Hz, 2Hx, 10 Hz,
70 Hz

Freq Range (Fmax)

The FFT upper full-scale
frequency, up to 40,000 Hz
(2,400,000 CPM).

1000 Hz (default)