Metrohm 700 Dosino User Manual

Page 31

Advertising
background image

3.1 Air bubbles are (almost) inevitable

700 Dosino, Instructions for Use

25

The standard assignment listed below applies to the following
Metrohm instruments; it cannot be altered:

726 / 796 Titroprocessor

736, 751 and 758 Titrino models

746 VA Trace Analyzer

756 KF-Coulometer

The standard occupancy of the Dosino ports:

Port 1

Dosing outlet; M6 threaded connection on the left-hand
side of the housing.
The liquid is ejected via dosing or titrating tip.

Port 2

Filling inlet; M6 threaded connection on the bottom of
the dosing unit.
The liquid is aspired from a storage container.

Port 3

Not assigned; M6 threaded connection on the right-
hand housing side.

Port 4

Special functions; slender connection nipple on the
bottom of the dosing unit.
This can be used with the

'PREP' function for ejecting

the liquid. When emptying the dosing unit port 4 is
used as the air inlet.

Port 0

Venting the storage bottle; M6 threaded connection at
the front. A drying tube can be connected here; this
can be filled with e.g. molecular sieve or soda lime.

Variable port assignments are possible with the following instru-
ments:

730 Sample changer (and successors)

711 Liquino

774 Oven Sample Processor

Ports 1 to 4 can be used bi-directionally here and are therefore
suitable for filling the dosing cylinder or dosing a liquid as required.
This means that complex liquid handling applications are possible.

If one of ports 1 to 3 is not in use then it should be closed
with a thread stopper (6.1446.040).
Never close Port 0 with a thread stopper when the dosing unit
is attached to a storage container. This could create a vac-
uum in the storage container – risk of implosion!