3.4 Troubleshooting / Problems

700 Dosino, Instructions for Use

33

The whole system is blocked.

Possible cause:

The Dosino or control instrument is in an exceptional
faulty condition.

⇒ Check cable connections.
⇒ Switch control instrument off and on again.

Remove dosing drive from dosing unit. With the instrument
switched on check whether the drive carrier can be rotated.
– If yes then the dosing drive is faulty.
– If no then dismantle the dosing unit. Clean the black stop-
cock washer in the cylinder base, see page 29.

3.4.2 Dosing unit

Air bubbles in cylinder or dosing tubing

Possible causes:

Leaking connections.

⇒ Check the tubing ends, in particular that of the aspiration tub-

ing.

⇒ Tighten all tubing connections manually.

Check whether the housing is properly locked. If necessary
remove the housing and then put it back on again.

The reagent releases a lot of gas, i.e. dissolved air forms
bubbles.

⇒ Carry out PREP.
⇒ Lower the filling speed.

If necessary degas the reagent ultrasonically.

Wear and tear.

Replace piston and/or cylinder.

PREP not carried out or incorrect parameters.

⇒ Carry out PREP.

Correct tubing lengths and diameters.

Stopcock washer and distributor disk stick together

Possible cause:

Precipitation of crystals from the reagent.

Clean the stopcock washer and distributor disk, see page 29.