2 dosing accuracy according to iso 8655-3, Vs cv 100 – Metrohm 700 Dosino User Manual

Page 44

Advertising
background image

4.2 Dosing accuracy according to ISO 8655-3

700 Dosino, Instructions for use

38

4.2 Dosing accuracy according to ISO 8655-3

Metrohm Ltd. guarantees the following limits (according to ISO
8655-3, draft) for Dosino burets with glass cylinders:

Max. permissible systematic error e

s

Nominal

cylinder volume

Maximum permissible systematic errors e

s

a)

in mL

rel.

abs.

2

± 0,3 %

± 6 µL

5

± 0,3 %

± 15 µL

10

± 0,2 %

± 20 µL

20

± 0,2 %

± 40 µL

50

± 0,2 %

± 100 µL

a)

Expressed as deviation of the mean of a tenfold measurement

from any given volume in the range of 10 to 100 % of the nominal
volume.

s

s

V

V

e

=

or in %

(

)

0

100

V

Vs

V

s

e

=

V

: mean

s

V

: selected volume

0

V

: nominal volume

Maximum permissible random error s

Nominal

cylinder volume

Maximum permissible random errors

repeatybility s

in mL

CV

b)

s

c)

2

± 0,1 %

± 2 µL

5

± 0,1 %

± 5 µL

10

± 0,07 %

± 7 µL

20

± 0,07 %

± 14 µL

50

± 0,05 %

± 25 µL

b)

Expressed as coefficient of variation of a tenfold measurement

from any given volume in the range of 10 to 100 % of the nominal
volume.

V

s

CV

100

=

c)

Expressed as repeatability standard deviation of a tenfold mea-

surement from any given volume in the range of 10 to 100 % of
the nominal volume.