Metrohm 700 Dosino User Manual

Page 42

Advertising
background image

3. The Dosino in practice

700 Dosino, Instructions for Use

36

Tubing not completely filled during PREP.

Possible cause:

Incorrect parameters have been defined for the dosing
unit.

Correct the tubing lengths and diameters; please refer to the
‘Instructions for Use’ of the control instrument.

Tubing not completely emptied during EMPTY.

Possible cause:

Incorrect parameters have been defined for the dosing
unit.

Correct the tubing lengths and diameters; please refer to the
‘Instructions for Use’ of the control instrument.

Wrong volume is dosed.

Possible cause:

Dosing unit mounted incorrectly or wrongly assembled.

⇒ Remove dosing unit and put it back on again.

Check whether the nominal volume on the housing coincides
with the effective cylinder volume.

Dosing does not take place.

Possible cause:

Blocked tubing connections or dosing unit wrongly as-
sembled.

⇒ Check whether the dosing tip is blocked.
⇒ Check whether the dosing port has been closed with a thread

stopper.

⇒ Check whether Port 0 has been closed with a thread stopper

(vacuum in storage bottle!). Port 0 must be open to compen-
sate the pressure.

Remove dosing drive and check whether the drive push rod
engages the dosing piston. The piston peg must be flush with
the upper side of the housing, see page 19.

Advertising