Vision reports that there is
no Reference
Standardization or
Instrument Calibration;
suggests that these tests be
run.

1. Usually caused by entering the test, then canceling,
which leaves the instrument with no completed test
constants. Do not cancel when these tests have been
initiated.

2. May be caused by use of two network cards in PC. Data
collisions are causing lost communication “packets”.
Remove one NIC card, and use the connection method on
page 4.

If these measures do not correct the problem, please contact your sales and distribution office.