2.3 Connection of the 825 Lab Link

Connect the 825 Lab Link to your switched-off Titrando system as
shown in Section 1.1 and connect it to the network:

• To do this connect the USB RS 232 Box to a USB connection

of the Titrando by using the 6.2151.030 USB Cable.

• Connect one RS 232 connection of the USB RS 232 Box to RS

232 connection 1 of the 825 Lab Link by using the 6.2134.040
Cable.

• Connect the 825 Lab Link to the local network by using an

RJ45 Ethernet cable at connection 3.

• For MSB connection 7 of the 825 Lab Link you can use any

free MSB connection of the Titrando or a peripheral device.
This connection is then made available again on the 825 Lab
Link to exactly the same extent (MSB connection 6).

2.4 Configuration of the Touch Control

The Touch Control used must be set for data transmission to an
external system:

1. Switch the Touch Control on.

2. If no PC/LIMS has been set up under System/Device

manager, then insert it with [New].

3. Select PC/LIMS and operate [Edit]. Set "Ext. System

access" as the communication mode.

4. You should also set the RS 232 interface according to

the selected connection at the USB RS 232 Box.

5. The Touch Control must then be switched off in order to store

these settings.