Metrohm 857 Titrando User Manual

Page 17

Advertising
background image

825 Lab Link, Instructions for Use

13

Advanced

Under Advanced special adaptations to the local network sur-
roundings can be carried out. These apply immediately for all Lab
Link connections after confirmation with [Ok].

Warning! You should only alter these settings when you have de-
tailed knowledge of the corresponding functions. Faulty timing
settings may cause malfunctioning.

Log-File level
Communication with the connected Titrando systems can be re-
corded in a log file. This function may be particularly useful during
startup for searching for errors. The log file is deleted every 60
operating minutes; recording then starts again from the begin-
ning.

Level low: No detailed recording is made. This is recommended
for normal operation.

Level medium: All received and transmitted commands as well as
the periodic status tests are recorded. This is normally sufficient
for tracing communication problems.

Level high: In addition, the info-packages sent by each 825 hard-
ware as well as technical programming information is recorded.