6.1

■■■■■■■■■■■■■■■■■■■■■■

94

■■■■■■■■

940 Professional IC Vario TWO/SeS (2.940.2400)

Problem

Cause

Remedy
Note: 6.2821.180 pump tubing connection
with filter must be used.

Guard column – Blocked.

Replace the guard column (see Chapter 3.20,
page 49)
.

Separation column –
Blocked.

Regenerate the separation column (see
Chapter 5.16.4, page 91)
.

Replace the separation column (see "Con-
necting the separation column", page 53)
.

Note: Samples should always be microfiltered .

Injection valve – Valve
blocked.

Have the valve cleaned (by a Metrohm service
technician).

The retention times
in the chromato-
gram have changed
unexpectedly.

Separation column –
Diminished separating effi-
ciency.

Regenerate the separation column (see
Chapter 5.16.4, page 91)
.

Replace the separation column (see "Con-
necting the separation column", page 53)
.

The eluent contains gas
bubbles.

Ensure that the connectors for the eluent
degasser are firmly connected .

Deaerate the high-pressure pump (see Fig-
ure , page 48)
.

High-pressure pump –
Defective.

Request Metrohm Service.

Peak areas are
lower than
expected.

Sample – There is a leak in
the sample path.

Find and fix the leak in the sample path.

Sample – The sample path
is blocked.

Find and fix the blockage in the sample path.

Sample – The sample loop
is not (completely) filled.

Adjust the transfer time for the sample (see
"Determining the transfer time", page 89)
.

Sample – There are gas
bubbles in the sample.

Use the sample degasser .

MCS – The CO

2

suppressor

is not connected.

Connect the CO

2

suppressor.

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean the contact surfaces of the column chip
with alcohol.

Column chip is defective.

1. Save column configuration in MagIC Net™.
2. Notify Metrohm Service.