Metrohm 940 Professional IC Vario TWO/SeS/PP User Manual

Page 94

Advertising
background image

5.12 Suppressor

■■■■■■■■■■■■■■■■■■■■■■

86

■■■■■■■■

940 Professional IC Vario TWO/SeS (2.940.2400)

Replacing parts of the suppressor

1 Disconnecting the suppressor from the IC system

Switch off the instrument.

Disconnect all capillaries of the suppressor from the IC system.

2 Dismantling the suppressor

Unscrew union nut (23-1) from the housing (23-4).

Pull the connecting piece (23-2) out of the housing together with
the the rotor (23-3).
If rotor A gets stuck in the housing, you can knock it loose as fol-
lows:
Put a pointed object into the slot in the housing and use it to
knock rotor A loose.

Detach the connecting piece from the rotor.

3 Cleaning the new rotor

Clean the surface of the new rotor (23-3) with ethanol using a
lint-free cloth.

4 Inserting the new rotor

CAUTION

Rotor A may be destroyed during start-up if not inserted correctly.

Insert the new rotor (23-3) into the housing (23-4) in such a way
that the tubing connections on the rear of rotor fit into the corre-
sponding recesses inside the housing and one of the three holes
of the rotor is visible from below in the slot of the housing (23-5).

Advertising