5 Operation and maintenance
940 Professional IC Vario TWO/SeS (2.940.2400)
■■■■■■■■
79
Inserting a new filter
1 Inserting a new filter
■
Place the new filter flat in the filter housing (22-1) using tweezers.
2 Mounting the filter screw
■
Screw the filter screw (22-2) back into the filter housing (22-1)
and tighten by hand. Then use two adjustable wrenches
(6.2621.000) to tighten it slightly.
3 Remounting the inline filter
■
Screw pressure screws back onto the inline filter.
4 Rinsing the inline filter
■
Dismantle the guard column (if present) and the separation col-
umn and replace with a coupling (6.2744.040).
■
Rinse the instrument with eluent.
■
Reinsert the columns after 10 minutes.