7 Technical specifications

940 Professional IC Vario TWO/SeS (2.940.2400)

■■■■■■■■

101

7.12

Detector

You can find the technical specifications for the detector in the detector's
manual.

7.13

Sample degasser

Material

Fluoropolymer

Resistance to sol-
vents

No restriction (except PFC)

Time to establish
vacuum

< 60 s

7.14

Power connection

Required voltage

100 - 240 V ± 10% (autosensing)

Required fre-
quency

50 - 60 Hz ± 3 Hz (autosensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

Internal fuse 3.15 A

7.15

Interfaces

USB

Input

1 USB upstream, type B (labeled PC for connecting to a computer)

Output

2 USB downstream, type A (labeled USB 1 and USB 2)

MSB

2 MSB 8-pin mini-DIN (female) (labeled MSB 1 and MSB 2 for Dosino,
stirrer, remote lines, etc.)

Detector

2 15-pin high-density D-sub (female) (labeled Detector 1 and Detec-
tor 2
)

Column recogni-
tion

3 (including 2 in the column thermostat )

Leak sensor

1 jack plug (labeled Leak Sensor)

Further connec-
tions

1 15-pin D-sub (female) (labeled Extension Module)