9 injection valve, 10 suppressor, 11 metrohm co2 suppressor (mcs) – Metrohm 940 Professional IC Vario TWO/SeS/PP User Manual
Page 108: Injection valve, 11 metrohm co, Suppressor (mcs)
Advertising
7.9 Injection valve
■■■■■■■■■■■■■■■■■■■■■■
100
■■■■■■■■
940 Professional IC Vario TWO/SeS (2.940.2400)
Gradient capabil-
ity
Isocratic or gradient (extendable to quaternary)
Profile
Step, linear, convex and concave
Resolution
< 1 nL/min
7.9
Injection valve
Actuator switch-
ing time
typ. 100 ms
Maximum operat-
ing pressure
35 MPa (350 bar)
Material
PEEK
7.10
Suppressor
Resistance to sol-
vents
No restriction
Switching time
typ. 100 ms
7.11
Metrohm CO
2
Suppressor (MCS)
Material
Fluoropolymer
Resistance to sol-
vents
No restriction (except PFC)
Underpressure
Working range
Microprocessor controlled/stabilized
Time to estab-
lish after start-
ing
< 30 s
Capillary volume
400 µL
Recommended
flow range
0.1 - 1.0 mL/min
Advertising