3.13 Installing the detector

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

940 Professional IC Vario ONE/Prep 2 (2.940.1120)

3.13

Installing the detector

The 940 Professional IC Vario provides enough space for two detectors
and additional accessories in the detector chamber. The detectors are
available as accessories and are supplied with separate manuals.

Placing the detector in the instrument
Follow the instructions in the chapter Inserting the detector in the manual
for the detector.

Connecting the detector to the eluent path

NOTE

The separation column is not inserted into the instrument until it is
being started up for the first time. Until then, the detector input capil-
lary has to be connected directly to the column input capillary using a
coupling (6.2744.040).

Accessories

For this step you need the following accessories:

Coupling (6.2744.040)

2 × pressure screw (6.2744.010)

1 Connect the column input capillary and the detector input capillary

to one another using a coupling (6.2744.040) and two short pres-
sure screws (6.2744.070).