3.15 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

940 Professional IC Vario ONE/Prep 2 (2.940.1120)

3 Inserting the pump tubing

Press the tubing cartridge's contact pressure lever down all the
way.

Place the pump tubing in the tubing cartridge. Fit the tubing car-
tridge between two stoppers. The stoppers must snap into the
corresponding holder of the tubing cartridge.

4 Inserting the tubing cartridge

Hang the tubing cartridge in the mounting bolt and press it in the
cartridge holder until you hear the snap-action lever snap in.

Setting the flow rate
The flow rate of the peristaltic pump depends on many factors:

The inner diameter of the pump tubing

The rotational speed of the drive

The contact pressure of the tubing cartridge

NOTE

Pieces of pump tubing are consumables. The lifetime of the pump tub-
ing depends on the contact pressure, among other factors.

Setting the contact pressure correctly

1

Fully loosen the contact pressure lever , i.e. press it all the way
down.

Activate the peristaltic pump drive in the software.

Raise the contact pressure lever one step at a time until liquid
flows.

When liquid starts flowing, raise the contact pressure lever by an
additional two ratchet increments.