Metrohm 940 Professional IC Vario ONE/Proep 2 User Manual
Page 43
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
940 Professional IC Vario ONE/Prep 2 (2.940.1120)
■■■■■■■■
35
1 Connecting the pump tubing
■
Attach the Coupling Olive/UNF 10/32 (6.2744.034) to the pump
tubing inlet. Push the end of the pump tubing over at least the
second notch of the olive so that the pump tubing is firmly in
place.
■
Install the pump tubing connection with safety device and filter
(6.2744.180) at the pump tubing outlet:
– Push the locknut onto the pump tubing.
– Push the appropriate adapter onto the pump tubing.
– Place the tubing olive with the filter holder into the pump
tubing so that the pump tubing is firmly in place; push the
end of the pump tubing over at least the second notch of
the olive.
– Tighten it using the union nut.
2 Removing the tubing cartridge
■
Press in the tubing cartridge's snap-action lever.
■
Tilt the tubing cartridge upwards.
■
Unhook the tubing cartridge from the mounting bolt.