6 Troubleshooting

940 Professional IC Vario ONE/Prep 2 (2.940.1120)

■■■■■■■■

81

Problem

Cause

Remedy

Injection valve – Valve
blocked.

Have the valve cleaned (by a Metrohm service
technician).

The retention times
in the chromato-
gram have changed
unexpectedly.

Separation column –
Diminished separating effi-
ciency.

Regenerate the separation column (see
Chapter 5.15.4, page 78)
.

Replace the separation column (see "Con-
necting the separation column", page 47)
.

The eluent contains gas
bubbles.

Ensure that the connectors for the eluent
degasser are firmly connected .

Deaerate the high-pressure pump (see Fig-
ure , page 42)
.

High-pressure pump –
Defective.

Request Metrohm Service.

Peak areas are
lower than
expected.

Sample – There is a leak in
the sample path.

Find and fix the leak in the sample path.

Sample – The sample path
is blocked.

Find and fix the blockage in the sample path.

Sample – The sample loop
is not (completely) filled.

Adjust the transfer time for the sample (see
"Determining the transfer time", page 74)
.

Sample – There are gas
bubbles in the sample.

Use the sample degasser .

The peristaltic pump
is pumping too little.

Peristaltic pump – Contact
pressure too weak.

Correctly set the contact pressure .

Peristaltic pump – Filter
blocked.

Replace the filter (see "Replacing the filter",
page 77)
.

Peristaltic pump – Pump
tubing defective.

Replace the pump tubing .

Data of the separa-
tion column cannot
be read.

Column chip contamina-
ted.

Clean the contact surfaces of the column chip
with alcohol.

Column chip is defective.

1. Save column configuration in MagIC Net™.
2. Notify Metrohm Service.

Individual peaks are
greater than
expected.

Sample – Sample carry-
over from the previous
measurement.

Check the rinsing time (see "Checking the rins-
ing time", page 75)
.