3.16 Installing the 6-port valve

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

940 Professional IC Vario ONE/Prep 2 (2.940.1120)

3.16

Installing the 6-port valve

During sample preparation, the 6-port valve is used as a switch between
two different flows.

Connectors
The 6-port valve (like the injection valve) has six connections that can be
connected together depending on the application.

Connecting the 6-port valve

1 Connect all capillaries using PEEK pressure screws (6.2744.010).

Valve positions
The 6-port valve (see Figure 14, page 38) can have two valve positions:
Filling and Injecting. Switching is used to connect two different connec-
tions together in each case.

Figure 14

6-port valve – Positions

1

Filling

2

Injecting

Filling

Ports 2 and 3, 4 and 5 as well as 6 and 1 are
connected together in the Filling position.

Injecting

Ports 1 and 2, 3 and 4 as well as 5 and 6 are
connected together in the Injecting position.