10 metrohm co2 suppressor (mcs), 11 mains connection, 12 interfaces – Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 108: 10 metrohm co, Suppressor (mcs)

Advertising
background image

7.10 Metrohm CO

2

Suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

98

■■■■■■■■

882 Compact IC plus – Anion – MCS

7.10

Metrohm CO

2

Suppressor (MCS)

Material

fluoropolymer

Resistance to sol-
vents

No restriction (apart from PFC)

Vacuum

Working area

Microprocessor-controlled / stabilized

Build-up time
after start

< 30 s

Capillary volume

400 µL

Recommended
flow range

0.1…1.0 mL

7.11

Mains connection

Required voltage

100…240 V ± 10 % (auto-sensing)

Required fre-
quency

50…60 Hz ± 3 (auto-sensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

internal fuse 3.15 A

7.12

Interfaces

USB

Input

1 USB upstream, type B (for connection to the PC)

Output

2 USB downstream, type A

MSB

2 MSB MiniDin 8-pin (female) (for Dosino, stirrer, remote lines, …)

Caution

When connecting an instrument to the MSB connector you must
switch off the 882 Compact IC plus.

Detector

1 DSUB-15-pin high density (female)

Column recogni-
tion

for an intelligent column

Advertising