3 installing the adsorption cartridges, Installing the adsorption cartridges, Figure 26 – Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 60: Adsorption cartridge holder

Advertising
background image

3.15 Metrohm CO

2

Suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

882 Compact IC plus – Anion – MCS

3.15.3

Installing the adsorption cartridges

For effective CO

2

removal, the air sucked through the degassing cell

should be as low in CO

2

as possible. To achieve this, the air is aspirated

through a CO

2

adsorption cartridge (6.2837.000) (26-4)..

Moisture can block the CO

2

adsorption cartridge. In order to prevent this,

a H

2

O adsorption cartridge (6.2837.010) (26-7) is connected upstream.

9

3

8

7

3

1

2

3

3

4

5

6

Figure 26

Adsorption cartridge holder

1

PEEK pressure screw, short
(6.2744.070)

2

Luer coupling (6.2744.120)

3

Clips
For fastening the adsorption cartridges.

4

CO

2

adsorption cartridge (6.2837.000)

For removing the CO

2

from the aspirated air.

3-layer filled, blue-brown-gray.

5

Adapter (6.1808.190)
For connecting the H

2

O adsorption cartridge

and CO

2

adsorption cartridge.

6

PVC tubing
For connecting the H

2

O adsorption cartridge

and CO

2

adsorption cartridge.

Advertising