3 installing the adsorption cartridges, Installing the adsorption cartridges, Figure 26 – Metrohm 882 Compact IC plus – Anion – MCS User Manual
Page 60: Adsorption cartridge holder
3.15 Metrohm CO
2
Suppressor (MCS)
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
882 Compact IC plus – Anion – MCS
3.15.3
Installing the adsorption cartridges
For effective CO
2
removal, the air sucked through the degassing cell
should be as low in CO
2
as possible. To achieve this, the air is aspirated
through a CO
2
adsorption cartridge (6.2837.000) (26-4)..
Moisture can block the CO
2
adsorption cartridge. In order to prevent this,
a H
2
O adsorption cartridge (6.2837.010) (26-7) is connected upstream.
9
3
8
7
3
1
2
3
3
4
5
6
Figure 26
Adsorption cartridge holder
1
PEEK pressure screw, short
(6.2744.070)
2
Luer coupling (6.2744.120)
3
Clips
For fastening the adsorption cartridges.
4
CO
2
adsorption cartridge (6.2837.000)
For removing the CO
2
from the aspirated air.
3-layer filled, blue-brown-gray.
5
Adapter (6.1808.190)
For connecting the H
2
O adsorption cartridge
and CO
2
adsorption cartridge.
6
PVC tubing
For connecting the H
2
O adsorption cartridge