3.12 Injection valve

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

882 Compact IC plus – Anion – MCS

2 Mounting new sample loop

Fasten one end of the sample loop (18-2) with a 6.2744.010
PEEK pressure screw (18-7) to connector 3.

Fasten the other end of the sample loop (18-2) with a second
6.2744.010 PEEK pressure screw (18-7) to connector 6.

3.12.2

Mode of operation of the injection valve

The injection valve (see Figure 19, page 38) can adopt two valve posi-
tions - FILL and INJECT. Switching back and forth between the two valve
positions determines whether the sample path or the eluent path is gui-
ded through the sample loop. The following figure provides a schematic
display of the flow paths of the two valve positions.

1

2

4

5

3

6

1

2

4

5

3

6

1

2

4

3

5

4

3

1

2

A

B

Figure 19

Injection valve – Positions

A

Position FILL

B

Position INJECT

1

Eluent input
Capillary coming from the high pressure
pump.

2

Eluent output
Capillary to the column.

3

Sample input
Sample aspirating capillary.

4

Sample output
Capillary to waste container.

5

Sample loop

Position A

In the position FILL, the sample solution flows
through the sample loop to the waste container.
The eluent flows directly to the separation col-
umn at the same time.