11 pulsation absorber, Figure 16, Connecting the inline filter – Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 45

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

882 Compact IC plus – Anion – MCS

■■■■■■■■

35

Installing the inline filter

Caution

Observe the flow direction marked on the filter housing for the connec-
tion of the inline filter.

2

3

2

4

1

Figure 16

Connecting the inline filter

1

Connection capillary
Connects the purge valve with the inline fil-
ter

2

PEEK pressure screws, short
(6.2744.070)

3

Inline filter (6.2821.120)
Protects against particles.

4

Connection capillary
Connects the inline filter with the pulsation
absorber.

1 Screw on the connection capillary running from the purge valve to

the input side of the inline filter using a pressure screw (6.2744.070).

2 Screw on the connection capillary running to the pulsation absorber

to the output side of the inline filter using a pressure screw
(6.2744.070).

3.11

Pulsation absorber

Note

The pulsation absorber is already installed in the newly delivered instru-
ment.

Caution

The pulsation absorber is maintenance-free and may not be opened.

Advertising