8 eluent, 1 connecting eluent bottle, Eluent – Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 35: Connecting eluent bottle

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

882 Compact IC plus – Anion – MCS

■■■■■■■■

25

3.8

Eluent

3.8.1

Connecting eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing (8-1).

The eluent aspiration tubing is connected to the high pressure pump (see
Chapter 3.9, page 29)
(see Chapter , page ). The tubing must be threa-
ded through a suitable capillary feed-through (see Chapter 3.7, page 23)
of the instrument before the other end can be equipped.

You will require the parts from the following accessories for equipping the
eluent aspiration tubing:

6.1602.160 Eluent bottle attachment GL 45

6.2744.210 tubing adapter for aspiration filter

6.2821.090 aspiration filter

To equip the eluent aspiration tubing proceed as follows:

Assembling eluent aspiration tubing

1 Guide the free end of the eluent aspiration tubing (8-1) out of the

instrument through a suitable capillary feed-through.

2 Installing eluent bottle attachment (6.1602.160)

Slide tubing nipple (8-2) and O-ring (8-3) onto the eluent aspira-
tion tubing (8-1).

Push eluent aspiration tubing (8-1) through the bottle attachment
(8-4) and screw tight.