Figure 8, Installing tubing weighting and aspiration filter – Metrohm 930 Compact IC Flex SeS User Manual

Page 29

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

930 Compact IC Flex SeS (2.930.1400)

■■■■■■■■

21

Place the loose end of the eluent aspiration tubing into the aspira-
tion filter.
The end of the tubing should reach approximately to the center of
the aspiration filter.

Tighten the aspiration filter to the filter holder.

Figure 8

Installing tubing weighting and aspiration filter

3 Installing the eluent bottle cap on the eluent bottle

Insert the eluent aspiration tubing into the eluent bottle
(6.1608.070).

Tighten the bottle cap on the eluent bottle.

Adjust the length of the eluent aspiration tubing so that the aspi-
ration filter is at the bottom of the eluent bottle. Then fasten it in
place using the M8 tubing nipple.

Seal the M6 opening on the bottle cap with the M6 threaded
stopper from the accessory set.