5.6 Servicing the high-pressure pump
■■■■■■■■■■■■■■■■■■■■■■
64
■■■■■■■■
930 Compact IC Flex SeS (2.930.1400)
■
Has the pressure dissipated?
■
Is the instrument switched off?
1 Release the coupling from the pressure screw and seal it with a stop-
per.
2 Unscrew and remove the pressure screw on the pump head's output
3 Unscrew and remove the four fastening screws (15-3) using the hex
key.
4 Remove the pump head (15-4).
Replacing the piston seal
Carry out the following work on both pistons in turn.
Servicing a piston consists of the following tasks:
1.
Replace the piston seal.
2.
Optional: Clean or replace the zirconium oxide piston.
3.
Reinstall the piston.