Figure 21, Inline filter – removing the filter – Metrohm 930 Compact IC Flex SeS User Manual
Page 80
Advertising
5.7 Servicing the inline filter
■■■■■■■■■■■■■■■■■■■■■■
72
■■■■■■■■
930 Compact IC Flex SeS (2.930.1400)
Removing the filter
Figure 21
Inline filter – removing the filter
1
Filter housing
Inline filter housing. Part of the accessories
(6.2821.120).
2
Filter screw
Screw for the inline filter. Part of the acces-
sories (6.2821.120).
3
Filter (6.2821.130)
Packaging contains 10 pieces.
1 Shutting off the flow
Switch off the high-pressure pump in the software.
2 Removing the inline filter
Unscrew the both pressure screws from the inline filter.
3 Unscrewing the filter screw
Use two adjustable wrenches (6.2621.000) to loosen the filter screw
(21-2) from the filter housing (21-1) and unscrew it by hand.
4 Removing the filter
Remove the old filter (21-3) using tweezers.