3.10 Suppressor

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

930 Compact IC Flex SeS (2.930.1400)

1 Fasten the capillary labeled regenerant to the outlet of the inline filter

(6.2821.120) using a pressure screw (6.2744.070).

2 Use a pressure screw (6.2744.070) to fasten one end of the PEEK

capillary (6.1803.030) to the inlet of the inline filter.

3 Use a pressure screw (6.2744.070) and an adapter (6.2744.080) to

fasten the other end of the PEEK capillary to Port 1 of the Dosino.

4 Connect the FEP tubing (6.1805.120) from the bottle with the regen-

eration solution to Port 2 of the Dosino.

3.10.2.3

Connecting the rinsing solution

There are 2 options for rinsing the suppressor:

Rinsing solution via STREAM (recommended)
Use the eluent from the conductivity detector as a rinsing solution.

Rinsing solution via Dosino
Prepare the rinsing solution in its own bottle and transport with a Dos-
ino.

The rinsing solution is connected to the rinsing solution capillary.