background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

909 UV Digester

■■■■■■■■

11

1

2

Figure 5

Installing the UV lamp in the wet end

3.3

Mounting the UV protective shield

A UV protective shield (6.2745.100) pre-installed on a metal ring is inclu-
ded in the scope of delivery for the 909 UV Digester.

Proceed as follows:

1 Using the hex key (6.2621.100), unscrew the stop's hex socket head

screw (2-6) on the back of the housing and remove the stop.

Keep the stop with the hex socket head screw in reach. It will have
to be screwed back in place on the housing afterwards.

Figure 6

Removing the stop

2 Place the UV protective shield over the wet end's opening so that the

metal ring rests on the two front stops (2-3).