2 care, 1 maintaining and replacing the uv lamp, Care – Metrohm 909 UV Digester User Manual

Page 42: Maintaining and replacing the uv lamp

Advertising
background image

7.2 Care

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

909 UV Digester

7.2

Care

7.2.1

Maintaining and replacing the UV lamp

Keep the following information in mind so that the lifetime of the UV
lamp is not shortened unnecessarily:

Clean the UV lamp as needed using a clean towel wetted with pure
alcohol. This lets you remove particles from the quartz glass that have
not yet been burned in.
Do not touch any part of the UV lamp other than the two flattened
ends during cleaning and wear the provided cotton glove when remov-
ing or installing the lamp.

Avoid switching the UV lamp on and off unnecessarily. Each time the
lamp is switched on shortens its lifetime.

Removing the old UV lamp

Warning

Risk of electric shock due to defective electronic components

Coming into contact with potentially defective electronic components
in the instrument could result in an electric shock.

Ensure that the instrument is unplugged before replacing the UV lamp.

Proceed as follows:

1 Ensure that the wet end has cooled completely.

2 Remove the sample holder from the wet end if it has been inserted.

3 Put on the glove included with the UV lamp.

4 Hold the UV lamp by the upper flattened end and gently press it

down.

5 While holding the UV lamp down, release the upper end from the

lamp socket.

6 Take the UV lamp out of the lower lamp socket and remove it from

the wet end.

Advertising