4 Operation

909 UV Digester

■■■■■■■■

21

The sample setpoint temperature (Sample Temp.).

The radiating time (Rad. Time).

Sample Temp.

Setpoint temperature for the samples.

Input range

80 - 100 °C (Increment: 1)

Default value

90 °C

Rad. Time

Radiating time for the samples.

Input range

1 - 999 min (Increment: 1)

Default value

90 min

Note

The configured parameters can be modified during digestion.

The configured parameters remain stored in the instrument after it is
switched off.

4.2.3

"Lamp info" dialog

You can open the Lamp info dialog from the main dialog by pressing the
[OK] key twice.

The following information is displayed:

The cumulative number of operating hours for the current UV lamp in
use since the last replacement (Hours).

The number of times the current UV lamp in use has been switched on
since the last replacement (Ignitions).

4.2.4

"Lamp setup" dialog