7 Operation and maintenance

909 UV Digester

■■■■■■■■

35

Inserting the new UV lamp

1 Insert the new UV lamp in accordance with Chapter 3.2, page 9.

Configuring parameters for the new UV lamp

1 Reset the counters to zero for the number of operating hours and

the number of times the lamp has been switched on under Lamp
setup

Reset Counters (see "Reset counters", page 22).

2 Set the number of operating hours for the new UV lamp under

Lamp setup

Replace after (see "Replace after", page 22).

We recommend 750 h of operating time.

7.2.2

Clearing a jammed fan

The wet end with inserted sample holder is designed so that no parts such
as PTFE stoppers can fall down. If the fan happens to get jammed anyway,
keep the following safety instruction in mind:

Warning

Jammed fan

Improper work on the fan may result in injuries.

Ensure that the instrument is unplugged before clearing the jammed
fan.

Proceed as follows to remove PTFE stoppers that have fallen into the wet
end of the 909 UV Digester:

1 Remove the sample holder.

2 Turn the instrument upside down.

3 Gently shake the overturned instrument until the PTFE stoppers come

out.