5 installing the flow-through cell, Installing the flow-through cell, Figure 5 – Metrohm 887 Professional UV/VIS Detector User Manual

Page 25: Cell block

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

887 Professional UV/VIS Detector

■■■■■■■■

17

2 Plug the capillary all the way into the connection or coupling until

the stop.

3 Only then start turning the pressure screw, while keeping the capil-

lary pressed in space.

Colored sleeves for PEEK capillaries
The enclosed set of varicolored sleeves for PEEK capillaries (6.2251.000)
serves to easily differentiate the various flows of liquid in the system
through color coding. Each capillary leading a given liquid (e. g. eluent)
can be highlighted with sleeves of the same color.

To highlight a capillary, proceed as follows:

1 Slide a sleeve of a selected color over a capillary an move it to an

easily visible position.

If the capillary heats up, the sleeve shrinks and adapts to the form of
the capillary.

3.5

Installing the flow-through cell

1

2

3

3

4

Figure 5

Cell block

1

Cell holder
Holder for the flow-through cell.

2

Covering plate
Protects the cell holder form contamination
if no cell is installed.

3

Knurled screws

4

Cylinder screw
For correct alignment of the flow-through
cell