8.3 Lamps

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

887 Professional UV/VIS Detector

Analog output

Voltage range

–1.0…+1.0 V

Resolution

30 µV

Noise

< 30 µV

Output impe-
dance

50

Ω (continuous short circuit fixed)

Offset error

± 1.5 mV

Plug

1 DSUB plug 9-pin (female)

1

2

3

4

5

6

7

8

9

0V

0V

0V

0V

0V

A4 (CH4) A3 (CH3) A2 (CH2) A1 (CH1)

Control input

Input voltage
range

0…5 V (5 V-Logical- or switching contact control possible.)

Impedance

22 k

Ω (permanent resistance up to 50 V. Secured against ESD.)

Function

Start, Zero, 2x Reserve
Caution: If input "reserve 2" is active at the time the instrument is
switched on, the instrument enters long-term test mode, which
switches through all elements following a time schedule and cannot be
terminated, not even by power failure. Can only be operated by service
software or RS-232 commands.

8.3

Lamps

UV lamp

Type

D2 (Deuterium)

Power con-
sumption

approx. 20…30 W

Lifetime

ca. 1000 h

VIS lamp

Type

Halogen

Power con-
sumption

approx. 5 W

Lifetime

approx. 10000 h