8 Technical specifications

887 Professional UV/VIS Detector

■■■■■■■■

37

8 Technical specifications

8.1

Reference conditions

The technical specifications listed in this chapter refer to the following ref-
erence conditions:

Ambient tempera-
ture

+25 °C (± 3 °C)

Instrument status

> 30 min in operation with both lamps switched on.

8.2

UV/VIS detector

Measuring range

Absorbtion

–2.0…+2.0 AU

Number of
channels

8 measuring channels, 4 of them readable in analog mode

Resolution

2.5 x 10

-7

AU

Noise

2.5 x 10

-5

AU (at data rate 1/s)

Drift

at reference
conditions

5 x 10

-4

AU/h

Wavelength range

wavelength
range

λ

190…900 nm

Wavelength
bunching

Δλ

1…limit nm (max. 101)
limit is 2 times the maximum number nm between the wavelength of
the corresponding channel and the nearest wavelength border.

Δλ is

always odd, since symmetrical around the wavelength of the corre-
sponding channel including the mean value.

Accuracy abso-
lute

± 3 nm

Stability

± 1 nm (over temperature range)

Optical resolu-
tion

5 nm (at 254 nm)

Measuring interval

Data rate for
each channel

0.5…50 Samples/sec