3 Installation

887 Professional UV/VIS Detector

■■■■■■■■

19

3.6

Connecting the flow-through cell

To connect the capillaries to the flow-through cell, proceed as follows:

Connecting the capillaries

1 Connecting detector input

Unscrew the pressure screw form the detector inlet IN.

Slide the pressure screw over the detector inlet capillary such that
a small part of the capillary is visible at the top.

Screw the detector inlet capillary to the detector inlet with the
pressure screw.

2

1

3

Figure 6

Connecting detector input

1

Detector inlet capillary
PEEK capillary (6.1831.100)

2

Detector inlet
Labeled with IN.

3

Pressure screw

2 Connecting detector output

Unscrew the pressure screw form the detector outlet OUT.

Slide the pressure screw over the detector inlet capillary such that
a small part of the capillary is visible at the top.

Screw the detector outlet capillary to the detector outlet with the
pressure screw.