Weller WR2 User Manual

Page 36

Advertising
background image

16-20

WR 2

Executing the calibration function (Factory Calibration Check)
With the FCC function you can check the temperature precision of

the repair station and even out possible deviations. For this purpose,

the soldering-tip temperature must be measured with an external

temperature meter and a temperature measuring tip assigned to the

soldering tool. The corresponding channel must be selected prior to

calibration.
Changing calibration at 100 °C / 210 °F
1. Insert the temperature sensor (0.5 mm) of the external

temperature meter into the temperature measuring tip.

2. Select the menu item FCC in Menu 2.

3. Press the DOWN button.

Calibration point 100 °C / 210 °F is selected.

The soldering tip is now heated to 100 °C / 210 °F.

The control indicator flashes as soon as the temperature is

constant.

4. Compare the temperatures indicated by the meter with the

indications in the display.

5. Use the UP or DOWN button to set the difference between the

value indicated on the external meter and the value indicated on

the repair station.

Maximum possible temperature adjustment ± 40 °C (± 70 °F).

Example:

Display 100 °C, external meter 98 °C: setting S 2

Display 100 °C, external meter 102 °C: setting T 2

Note Press button ┌ 2 ┐ to exit the menu item without changes (EXIT).

6. Press the ┌1·2 ┐ (Set) button to confirm the value.

The temperature deviation is now reset to 0. Calibration at 100 °C

/ 210 °F is now concluded.

7. Exit Menu 2 with button ┌ 2 ┐.
Changing calibration at 450 °C / 840 °F
1. Insert the temperature sensor (0.5 mm) of the external

temperature meter into the temperature measuring tip.

2. Select the menu item FCC in Menu 2.

3. Press the UP button.

Calibration point 450 °C / 840 °F is selected.

The soldering tip is now heated to 450 °C / 840 °F.

The control indicator (21) flashes as soon as the temperature is

constant.

4. Compare the temperatures indicated by the meter with the

indications in the display.

5. Use the UP or DOWN button to set the difference between the

value indicated on the external meter and the value indicated on

the repair station.

Maximum possible temperature adjustment ± 40 °C (± 70 °F).

Example:

Display 450 °C, external meter 448 °C: setting S 2

Display 450 °C, external meter 452 °C: setting T 2

Note Press button ┌ 2 ┐ to exit the menu item without changes (EXIT).

6. Press the ┌1·2 ┐ (Set) button to confirm the value.

The temperature deviation is now reset to 0. Calibration at 450 °C

/ 840 °F is now concluded.

7. Exit Menu 2 with button ┌ 2 ┐.