background image

35

35

Fig. 16. Selecting a native electrophoresis buffer system.

Determine the pI of the protein

under investigation. Isoelectric

focusing is recommended for

this purpose.

Determine the pH

range where the

protein retains its

biological activity.

NO

Ornstein-Davis buffer system

(pH 9.5) should not be used

because:

1. The protein of interest will

denature due to high pH.

2. Or, the protein of interest

will carry so little net charge

that it will not migrate

through the gel.

3. Or, the protein of interest

will carry a positive charge.

Is the pI of the pro-
tein below pH 8.5?

YES

NO

YES

NO

YES

Is the protein stable at

pH 9.5?

Does the protein precipitate

at the interface of the

stacking gel and resolving

gel?

Ornstein-Davis system can

be used. For reagent

preparation and running

conditions refer to Section 3.

A. Use a different

discontinuous system of

appropriate pH. (See

Chrambach and Jovin

1984)

B. Select a continuous buffer

system (McLellan1982) of

appropriate pH. See

Section 4.

Run the sample containing

the protein to be purified on

mini-slab gels using the

discontinuous system of

Ornstein and Davis. See

Section 3.

or