Bio-Rad Model 491 Prep Cell and Mini Prep Cell User Manual

Page 7

Advertising
background image

2

1.3 Specifications

Construction

Upper buffer chamber

acrylic

Lower buffer chamber

acrylic

Electrodes

platinum, 0.010 inch diameter

Lid acrylic
Gel tube assembly

glass/acrylic

Elution chamber base

acrylic

Elution frit

polyethylene

Support frit

polyethylene

Cooling core

glazed alumina

Elution tube

borosilicate glass, 0.06” ID

Casting stand

acrylic

Shipping weight

6 lb

Overall size

7 in. diameter x 14 in. high

Voltage limit

500 volts

Current limit

40 milliamperes

Power limit

20 watts

Cooling buffer flow rate

100 ml/min

Elution buffer flow rate

60 ml/hour

Upper electrophoresis buffer volume

300–600 ml

Elution buffer chamber volume

900 ml

Lower electrophoresis buffer volume

2–3 L

1.4 Chemical Compatibility

The Model 491 prep cell is not compatible with chlorinated hydrocarbons (e.g. chlo roform), aromatic
hydrocarbons (e.g. toluene, ben zene), or acetone. Their use will void all warranties.

1.5 Safety

Power to the Model 491 prep cell is to be supplied by an external DC power supply. This power
supply must be ground isolated in such a way that the DC voltage out put floats with respect to
ground. The recommended power sup ply for this instru ment is the PowerPac HV power supply.
The PowerPac universal power supply may also be used. The maximum speci fied operating
parameters for the Model 491 prep cell are:

500 VDC

maximum operating voltage

40 mA

maximum operating current

20 W

maximum operating power limit

Current to the Model 491 prep cell, provided from the external power supply, enters the unit
through the lid assembly, providing a safety interlock to the user. Current flow to the cell is
broken when the lid assembly is removed. Do not attempt to cir cumvent this safety interlock,
and always turn the power supply off when working with the cell.