Run assay, Considerations, Considerations when using a vacuum manifold – Bio-Rad Bio-Plex Pro™ Human Chemokine Assays User Manual

Page 20

Advertising
background image

18

8. Run Assay

Considerations

n

Bring all assay components and samples to room temperature

before use

n

Use calibrated pipets and pipet carefully, avoiding bubbles

n

Pay close attention to vortexing, shaking, and incubation instructions.

Deviation from the protocol may result in low assay signal and
assay variability

n

Assay incubations are carried out on a shaker at 850 ± 50 rpm at room

temperature (RT). Cover the plate with sealing tape and protect from
light with aluminum foil

Table 8. Summary of wash options and protocols. After each assay step, select the

appropriate Bio-Plex Pro

wash station program or perform the appropriate manual wash

step as summarized below.

Considerations When Using a Vacuum Manifold

n

After each incubation, place the filter plate on a calibrated vacuum

apparatus and remove the liquid by vacuum filtration

n

To wash, add 100 μl wash buffer to each well and remove the liquid as

before. Ensure that all wells are exposed to the vacuum

n

Thoroughly blot the bottom of the filter plate with a clean paper towel

between each vacuum step to prevent cross contamination

n

Place the assay plate on the plastic plate holder/tray as needed

n

Before each incubation, gently cover the plate with a new sheet of

sealing tape. Avoid pressing down on the wells to prevent leaking from
the bottom

Bio-Plex Pro or

Bio-Plex Pro II

Handheld Magnet or

Pro II Wash Station

Wash Station

Vacuum Manifold

Assay Step

Magnetic Program

Vacuum Program

Manual Wash Steps

Add beads to plate

MAG x2

VAC x2

2 x 100 μl

Sample incubation
Detection Ab incubation

MAG x3

VAC x3

3 x 100 μl

SA-PE incubation

Advertising