Instrument controls – Bio-Rad Microplate Manager Software User Manual

Page 19

Advertising
background image

Quick Start

11

Open a saved protocol or set up a new protocol (by setting up a template, and
selecting the instrument, assay info, analysis options and print options). Refer to
the

Instrument Controls

1. Click File > Read New Plate. The Instrument Controls dialog displays.

2. Under the Reader tab, set the Reading mode: Endpoint, Kinetic, or

Plate 2-1.

If you are using an xMark, Spectra is another Reading mode option.

3. Select the other Reading options:

Read speed (iMark only)
• Wavelength (nm) and single / dual
• Mix time & speed, and mix type (xMark only)
• Pathlength Correction

1

1

Pathlength correction is used to calculate the OD (or % transmission)

for each well, normalized to 1 cm fluid depth (d) in each well.