Technical features – Carlo Gavazzi WM3-96 User Manual

Page 61

Advertising
background image

62

Technical features

57

General specifications

64

56

64

Additional errors

DIGITAL FILTER:
• Operating range: 0 to 99.9% of the input scale
• Filtering coefficient: 1 to 255
• Filter action: alarms, serial and analogue outputs (fundamental

variables: V, A, W and their derived ones).

EVENT LOGGING: only with the module RS232+RTC mod-
ule. The max./min. values of the selected variables and the
alarm status are stored with reference to date (dd:mm:yy) and
time (hh:mm:ss). Max. capacity: 480 events.
VARIABLE PAGES: 4 variables per page; up to 27 pages, one
of which is completely programmable.

Harmonic Distortion Analysis

ANALYSIS PRINCIPLE: FFT
HARMONIC MEASUREMENT:
• Current: up to the 50

th

harmonic

• Voltage: up to the 50

th

harmonic

TYPE OF HARMONICS:
• THD (VL1); THD odd (VL1); THD even (VL1); the same for

the other phases: L2 and L3.

• THD (AL1); THD odd (AL1); THD even (AL1); the same for

the other phases: L2 and L3.

HARMONIC PHASE ANGLE:
• The instrument measures the angle between the single

harmonics of “V” and the single harmonics of “A” expressed
as “

°

”, making it possible to understand if the harmonics are

generated or imported.

On three-wire systems, the angle cannot be
measured.